Between planned maintenance visits

Protect rotor and sample safety.

A practical cross-platform routine for safer rotors, earlier warning signs and more reliable separations.

CENTRIFUGESLaboratory centrifuge housekeeping guide
Rotor condition is a safety-critical control.Correct assembly, balanced loading, prompt cleaning and visible condition checks protect people, samples and the centrifuge between planned engineering visits.

Take the routine into the laboratory

Choose a practical monthly task sheet for a physical instrument folder or an electronically completed laboratory record.

↓ Print-friendly PDF2-page A4 landscape · Rev 02
Low-ink version for printing
↓ Electronic record PDF2-page A4 landscape · Rev 02
Full-colour · complete and save electronically
Annual servicing is one important control—not a year-round maintenance strategy.
Laboratory centrifuges experience repeated high-speed loading, chemical exposure, temperature cycling and handling throughout the year. Their condition needs maintaining between service visits through correct loading, prompt cleaning, rotor inspection, usage-life control and timely response to vibration or damage.
IECIEC 61010-2-020:2026 ↗
Laboratory action

The laboratory’s role

Select the frequency that applies today and complete the checks shown.

⚠

Pause and investigate when you see:

Small changes can become preventable failures when they are allowed to persist.

•
Unusual vibration or imbalanceStop the run when vibration exceeds normal behaviour, the load shifts or an imbalance warning appears.
•
Rotor or bucket damageCracks, pitting, corrosion, deformation, abrasion or a dropped-component history require assessment before reuse.
•
Lid, lock or rotor-security concernThe lid or rotor will not secure normally, unlocks unexpectedly or shows damaged locking hardware.
•
Spill, leakage or tube breakageLiquid, salt deposits, glass or plastic fragments, aerosol release or unexplained residue is found in the rotor or chamber.
•
Speed or temperature deviationDisplayed or checked speed, time or temperature does not reach or maintain the expected condition.
•
Noise, odour or repeated errorsGrinding, rubbing, electrical odour or recurring drive, lid, imbalance or safety messages develop.
One care standard · three OEM perspectives

Apply the routine to your instrument

Start with the cross-platform routine, then use the emphasis relevant to your installed system.

Exact commands, parts, intervals and acceptance criteria remain model- and method-specific.

Evidence and source library

Safety, quality-system and OEM material supporting this cross-platform practical guide.

Primary sources
IEC 61010-2-020:2026 · Particular requirements for laboratory centrifugesCurrent fourth edition of the centrifuge-specific electrical equipment safety standard; use with the applicable edition of IEC 61010-1.
View official IEC record ↗
UK HSE · Provision and Use of Work Equipment Regulations 1998 (PUWER)Public overview covering suitable work equipment, inspection, maintenance, training and competence.
View HSE guidance ↗
Laboratory quality context · ISO 15189:2022 | ISO/IEC 17025:2017Apply where relevant to the laboratory. These standards support controlled equipment, competence and valid results; they do not prescribe this guide's exact housekeeping frequencies.
Reference standards

References are provided for transparency. IEC and ISO controlled current editions remain authoritative. PUWER applies within its UK scope; ISO 15189 and ISO/IEC 17025 apply according to the laboratory's activities and accreditation. OEM examples illustrate shared care principles but do not replace the exact centrifuge and rotor manuals. Always apply the validated method, risk assessment, biosafety or chemical controls and laboratory SOP for the installed configuration and use.

Lab Optimise practical laboratory guides

A growing cross-platform collection for laboratory care between planned maintenance visits.